CellOx 32517Operating manualDissolved oxygen sensor
Maintenance, cleaning, replacement CellOx 325265.4 Cleaning the electrodesCautionBefore starting to work with the sensor, please note the GENERAL MAIN
CellOx 325 Maintenance, cleaning, replacement27Thoroughly rinse the sensor head with deionized water.Water the counter electrode in deionized water fo
Maintenance, cleaning, replacement CellOx 325285.5 Checking the sensor for freedom from zero-currentThe sensor is zero-current free. Checking the sens
CellOx 325 What to do if...295.6 DisposalCautionThe ELY/G electrolyte solution irritates eyes and skin. When dealing with the ELY/G electrolyte soluti
Technical data CellOx 325307 Technical dataGeneral features Measuring principle Membrane covered galvanic sensorTemperature compensation IMT compensat
CellOx 325 Technical data31Pin assignmentMeasurementconditionsMeasuring ranges at 20 °C 0 ... 50 mg/l O20 ... 600 % O2 saturation0 ... 1250 mbar O2 pa
Wear parts and accessories CellOx 325328 Wear parts and accessoriesNoteFor further accessories, refer to the WTW catalog or the Internet.Wear parts an
CellOx 32518Copyright © Weilheim 2002, WTW GmbH & Co. KGReprinting - even as excerpts - is only allowed with the explicit written autho-rization o
CellOx 325 Contents19CellOx 325 - Contents1 Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .202 Safety . . . . .
Overview CellOx 325201 OverviewConstructionRecommendedfields ofapplicationl On site measurements in rivers, lakes and wastewaterl Applications in wate
CellOx 325 Safety212SafetyThis operating manual contains special instructions that must be followed during the operation of the Dissolved Oxygen (D.O.
Measuring / Operation CellOx 325224 Measuring / Operation4.1 CalibrationNoteFor calibration, please refer to the operating manual of the measuring ins
CellOx 325 Maintenance, cleaning, replacement235 Maintenance, cleaning, replacement5.1 General maintenance instructionsFor your safety When dealing wi
Maintenance, cleaning, replacement CellOx 325245.3 Changing the electrolyte and membrane headCautionBefore starting to work with the sensor, please no
CellOx 325 Maintenance, cleaning, replacement25Carefully shake off the drops of water.Fill a new membrane cap with ELY/G electrolyte solution. Remove