What to do if... pH/ 3400i
64
No stable measured
value
Obviously incorrect
measured values
Buffer solutions
Incorrect buffer solutions Change calibration procedure
Buffer solutions too old Use only once.
Note the shelf life
Buffer solutions depleted Change solutions
Cause Remedy
pH electrode:
Diaphragm contaminated Clean diaphragm
Membrane contaminated Clean membrane
Test sample:
pH value not stable Measure with air excluded if
necessary
Temperature not stable Adjust temperature if
necessary
Electrode + test sample:
Conductivity too low Use suitable electrode
Temperature too high Use suitable electrode
Organic liquids Use suitable electrode
Cause Remedy
pH electrode:
pH electrode unsuitable Use suitable electrode
Temperature difference
between buffer and test
sample too high
Adjust temperature of buffers
or test samples
Measurement procedure not
suitable
Follow special procedure