Ion concentration ProLab 4000
118
ba75646e02 01/2010
10 Wait for the measurement with stability control to end.
or
Take over the measured value with [OK].
The Standard register opens up.
11 Add standard solution to the sample (2 percent by volume of
the sample solution).
12 Start the measurement with [Continue].
The measured value is checked for stability (Stability control).
13 Wait for the measurement with stability control to end.
or
Take over the measured value with [OK].
The Calculation register with the calculated measured value
pops up.
14 Change the displayed unit of the measured value with [Unit] if
necessary.
15 If required, display the record of the current measurement with
[Record].
16 Start measuring further samples with [Continue].
Repeat the steps 2 - 14 for all samples.
or
Terminate the measuring method with [Cancel] and confirm the
safety query with [OK].
Measurement with the selected method is completed.