Multi 3400i Operation
55
4.7.2 PC/external printer (RS232 interface)
Via the RS 232 interface, you can transmit the data to a PC or an ex-
ternal printer.
Use the AK340/B (PC) or AK325/S (ext. printer) cable.
Warning
The RS232 interface is not galvanically isolated.
When connecting an earthed PC/printer, measurements cannot be per-
formed in earthed media as incorrect values would result.
Set up the following transmission data on the PC/printer:
Socket assignment
4.7.3 Remote control
The handheld meter can be remotely controlled from a PC. This re-
quires the KOM pilot communication kit. It is available as an accessory.
The instrument is then controlled via commands that simulate key-
strokes and request the current display contents.
Note
A more detailed description is provided within the scope of delivery of
the communication kit.
Baud rate can be selected from: 1200, 2400, 4800, 9600
Handshake RTS/CTS + Xon/Xoff
PC only:
Parity none
Data bits 8
Stop bits 2
1
2 RxD
3 Ground
4TxD
1
2
3
4