Multi 3400i What to do if...
67
6.2 Oxi system messages
Error message 0FL
0FL0FL
0FL
Error message E3
E3E3
E3
Error message E7
E7E7
E7
AR flashes
continuously
Cause Remedy
Display range exceeded
Oxygen sensor
– Not connected – Connect the probe
– Cable broken – Replace probe
– Depleted – Replace probe
– Short-circuit between gold and
lead electrode
– Clean probe and replace it if
necessary
Cause Remedy
Invalid calibration
Oxygen sensor
– Electrolyte solution depleted – Regenerate probe
– Membrane contaminated – Clean membrane
– Electrode system poisoned – Regenerate probe
– Worn out – Replace probe
– broken – Replace probe
Cause Remedy
Membrane damaged
– Membrane damaged – Regenerate probe
– Membrane head not screwed
on tight enough
– Screw membrane head tight
Cause Remedy
No stable measured value
– Membrane contaminated – Clean membrane