MultiLab 4010-2 What to do if...
ba76142e01 05/2013 91
13.3 Conductivity
Error message,
OFL
Error message,
Error
Measured temperature value outside
the operating conditions
(display of OFL/UFL instead of a
temperature value)
Keep to the temperature range
for the test sample
Defective sensor Calibrate
Change sensor cap
Replace sensor
More information and instructions on cleaning and exchange of
sensors are given in the documentation of your sensor.
Cause Remedy
Cause Remedy
Measured value outside the measur-
ing range
Use suitable IDS conductivity
sensor
Cause Remedy
Sensor contaminated Clean the sensor and replace it
if necessary
Calibration solution not suitable Check the calibration solutions