What to do if... MultiLab 4010-2
92 ba76142e01 05/2013
13.4 General information
Sensor symbol
flashes
Meter does not react
to keystroke
You want to know
which software
version is in the
meter or IDS sensor
Data transmission to
USB memory device
does not work
Cause Remedy
Calibration interval expired Recalibrate the measuring sys-
tem
Cause Remedy
Operating condition undefined or
EMC load unallowed
Processor reset:
Press the <ENTER> and <On/
Off> key simultaneously
Cause Remedy
E. g., a question by the service
department
Switch on the meter.
Open the menu, <ENTER_> /
Storage & config / System / Ser-
vice information. The instrument
data are displayed.
or
Connect the sensor.
Press softkey [Info]/[More] The
sensor data are displayed (see
section 4.1.6 S
ENSOR INFO,
page 17)
Cause Remedy
Connected USB memory device was
not recognized
The USB-B interface is connected to
a PC
The USB memory device has been
formatted to a file system which is
not supported, e. g. NTFS
Use other USB memory device
Disconnect the PC from the
USB-B interface
Reformat the USB memory
device to FAT 16 or FAT 32 file
system
(Caution
: Reformatting erases
all data on the USB memory
device. Back up all data before
reformatting.)