Operation Lab 850
34
ba75553e03 01/2006
4.5.2 Measurement settings
These settings apply to calibration and measurement (the default
setting is printed in bold).
Buffer set for pH
calibration
(SET)
Number of the buffer set for pH
calibration (SET)
1 ... 16
Display during calibration (BUFFER) pH (buffer nominal value),
U (electrode voltage)
Unit of the value for the zero point
(ASY)
pH, mV
Unit of the value for the slope (SLO) %, mV/pH
Temperature unit (UnIt) °C, °F
Calibration interval (Int.C) 0 ... 7 ... 999 d
1 Open the menu for measurement settings with <MODE_>.
Set 1 ... 16, the specified buffer set is displayed.
2 Using <><>, select a buffer set Set 1 ... 16.
3 Confirm with <OK>.
bUFF, the setting of the display during calibration is displayed.
1
AutoCal
SET
Seitenansicht 33
1 2 ... 29 30 31 32 33 34 35 36 37 38 39 ... 60 61