Lab 850 Operation
39
ba75553e03 01/2006
4.6.2 Resetting all meter settings
This function resets to the default condition all meter settings including
the calibration values. The corresponding values are given on the
following pages:
Resetting the meter
settings
Note
The measuring system is not calibrated after a reset. Before measuring
recalibrate the meter.
Calibration values page 38
System settings page 32
Measurement settings page 34
1 Switch on the meter with <On/Off>.
The display test appears briefly on the display.
2 During the display test, press <MODE> to open the menu for
the reset of the meter settings.
Init is displayed.
3 Press <><> to display no or YES.
YES: Reset the meter settings.
no: Retain the meter settings.
4 Confirm with <OK>.
The menu is finished.
The meter switches to the measuring mode.