Xylem Lab 850, up to serial number 09470314 Bedienungsanleitung Seite 43

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 61
  • Inhaltsverzeichnis
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 42
Lab 850 What to do if...
43
ba75553e03 01/2006
6 What to do if...
Error message
Err1
Error message
Err2
Error message
Err4
Error message
CalError
Cause Remedy
pH electrode:
Measured value outside the
measuring range
Use a suitable electrode
Air bubble in front of the
junction
Remove air bubble
Air in the junction Extract air or moisten
junction
Cable broken Replace the electrode
Gel electrolyte dried out Replace the electrode
Cause Remedy
No electrode connected Connect the electrode
Setting time during calibration
too long
Adjust temperature if
necessary
Recalibrate
Cause Remedy
Temperature not stable during
calibration.
Adjust temperature if
necessary
Recalibrate
Cause Remedy
pH electrode:
The values determined for zero
point and slope of the electrode
are outside the allowed limits.
Recalibrate
Junction contaminated Clean junction
Electrode broken Replace the electrode
Buffer solutions:
Seitenansicht 42
1 2 ... 38 39 40 41 42 43 44 45 46 47 48 ... 60 61

Kommentare zu diesen Handbüchern

Keine Kommentare