Ion concentration ProLab 3000
134
ba75645e03 12/2012
9 Immerse the electrode in the sample that was supplemented
with blank value solution.
10 Start the measurement with [Continue].
The measured value is checked for stability (Stability control).
11 Wait for the measurement with stability control to end.
or
Take over the measured value with [OK].
The Standard register opens up.
12 In the Standard register, enter the values for Standard volume
and Standard concentration if necessary.
13 Use [Dosing] to automatically add to the sample the adjusted
volume of the standard solution (Standard volume).
or
If the burette control is inactive:
Add to the sample the adjusted volume of the standard solution
manually (Standard volume).