ProLab 3000 Overview
19
ba75645e03 12/2012
1.5.2 Sensor data from ID sensors
ID sensors transmit the following sensor data to the meter:
SENSOR ID
Sensor type
Sensor series number
Calibration data
Calibration date
User (having carried out the last calibration procedure)
Calibration characteristics
Calibration interval
Selected buffer set (pH electrodes only)
The calibration data is updated in the ID sensor after each calibration
procedure. A message is displayed while data is updated in the sensor.
Note
While data is updated in the sensor, the sensor must not be discon-
nected because otherwise the calibration data will not be completely
transmitted. The sensor will then have no valid calibration.
Note
If non-ID sensors are used, the calibration data from the meter is used
and also stored in the meter.