7.3.3 Calibration settings and calibration data
In the Calibration menu, you can
set the calibration interval,
define the calibration type and
view the calibration data.
Menu item Setting Description
Calibration
pH calibration type
AutoCal () Calibration with commercial buffer sets.
The buffer set is selected with the next
menu item (Buffer sets)
Buffer sets DIN ()
TEC
...
Buffer sets to be used for pH calibration.
For details, see page 72.
VariCal () Calibration with any buffer solutions
Slope in mV/pH ()
%
Unit of the slope.
The % display refers to the Nernst slope of
-59.16 mV/pH ([determined slope/Nernst
slope ] x 100).
Zero point in mV ()
pH
Unit for the zero point.
Calibration interval... 1 ... 999 Calibration interval... for the pH electrode
(in days).
The meter reminds you to calibrate regu-
larly by the red frame around the CalClock
in the measuring window.
Calibration history... Displays the calibration history of the last
calibrations.